Prosecution Insights
Last updated: April 19, 2026
Application No. 17/977,671

DUAL LUMEN PATIENT LINE SYSTEM AND METHOD HAVING KINK DETECTION

Non-Final OA §103
Filed
Oct 31, 2022
Examiner
GERMAIN, ADAM ADRIEN
Art Unit
1777
Tech Center
1700 — Chemical & Materials Engineering
Assignee
BAXTER HEALTHCARE SA
OA Round
3 (Non-Final)
11%
Grant Probability
At Risk
3-4
OA Rounds
3y 2m
To Grant
-4%
With Interview

Examiner Intelligence

Grants only 11% of cases
11%
Career Allow Rate
3 granted / 27 resolved
-53.9% vs TC avg
Minimal -15% lift
Without
With
+-15.0%
Interview Lift
resolved cases with interview
Typical timeline
3y 2m
Avg Prosecution
79 currently pending
Career history
106
Total Applications
across all art units

Statute-Specific Performance

§101
4.2%
-35.8% vs TC avg
§103
54.2%
+14.2% vs TC avg
§102
14.4%
-25.6% vs TC avg
§112
25.4%
-14.6% vs TC avg
Black line = Tech Center average estimate • Based on career data from 27 resolved cases

Office Action

§103
DETAILED ACTION Notice of Pre-AIA or AIA Status The present application, filed on or after March 16, 2013, is being examined under the first inventor to file provisions of the AIA . Continued Examination Under 37 CFR 1.114 A request for continued examination under 37 CFR 1.114, including the fee set forth in 37 CFR 1.17(e), was filed in this application after final rejection. Since this application is eligible for continued examination under 37 CFR 1.114, and the fee set forth in 37 CFR 1.17(e) has been timely paid, the finality of the previous Office action has been withdrawn pursuant to 37 CFR 1.114. Applicant's submission filed on 16 JANUARY 2026 has been entered. Claim Status Rejected Claims: 1-4, 6-11, 13-18, and 20 Cancelled Claims: 5, 12, and 19 Response to Amendment The amendment filed on 16 JANUARY 2026 has been entered. In view of the amendment to the claims, the amendment of claims 1 and 16 has been acknowledged. In view of the amendment to claims 1 and 16, the rejections under 35 U.S.C. 112(b) have been withdrawn. In view of the amendment tot claims 1 and 16, the rejection under 35 U.S.C. 103 has been modified. Response to Arguments Applicant’s arguments filed on 16 JANUARY 2026 have been fully considered. Applicant argues, regarding claims 1 and 16, that Wallenas in view of Burbank does not explicitly teach the detecting an occlusion in a transfer set when both conditions (i) [an occlusion in the fill lumen] and (ii) [an occlusion in the return lumen] are met because Wallenas only teaches an occlusion detection during a forward or reverse flow of a pump and does not teach separate pressure sensors on each line (Arguments filed 16 JANUARY 2026, Pages 7-9). Regarding Applicant’s arguments, Wallenas teaches a continuous circulation of peritoneal fluid into and out of the peritoneal cavity in the second embodiment (Paragraphs 0073-0075) where pressure sensors are located in ports 73 and 86 (Fig. 2 below; Paragraph 0089) and describes kink detection if there is too high of pressure on pump with pressure sensor 87 in the cassette (Paragraph 0092) and that kink detection is important if sudden and unexpected deviation of pressure is measured (Paragraph 0094). PNG media_image1.png 354 734 media_image1.png Greyscale While not explicitly mentioning the transfer set, the alarm (iii) of the instant application is simply a derived alarm from a combination of alarms that are already triggered by Wallenas. Furthermore, Applicant’s arguments with respect to claims 1 and 16 have been considered but are moot because the new ground of rejection does not rely on any reference applied in the prior rejection of record for any teaching or matter specifically challenged in the argument. New art, of Mantle US Patent Application No. 20060161107 A1 (hereinafter Mantle), has been introduced which teaches the diagnostic procedure of testing pressure in multiple directions to determine where blockages are occurring in a double-barreled fluid line connected to a double-barreled catheter. Therefore, claim 1 is not allowable over Wallenas in view of Burbank in view of Mantle. Applicant argues, regarding claims 2-4, 6-11, 13-15, 17-18, and 20, that these claims depend upon claims 1 and 16, which are allowable, and so the dependent claims are also allowable (Arguments filed 16 JANUARY 2026, Page 9). Regarding Applicant’s arguments for the dependent claims, claims 1 and 16 are not allowable and so claims 2-4, 6-11, 13-15, 17-18, and 20 are also not allowable. Claim Rejections - 35 USC § 103 In the event the determination of the status of the application as subject to AIA 35 U.S.C. 102 and 103 (or as subject to pre-AIA 35 U.S.C. 102 and 103) is incorrect, any correction of the statutory basis (i.e., changing from AIA to pre-AIA ) for the rejection will not be considered a new ground of rejection if the prior art relied upon, and the rationale supporting the rejection, would be the same under either status. The following is a quotation of 35 U.S.C. 103 which forms the basis for all obviousness rejections set forth in this Office action: A patent for a claimed invention may not be obtained, notwithstanding that the claimed invention is not identically disclosed as set forth in section 102, if the differences between the claimed invention and the prior art are such that the claimed invention as a whole would have been obvious before the effective filing date of the claimed invention to a person having ordinary skill in the art to which the claimed invention pertains. Patentability shall not be negated by the manner in which the invention was made. The factual inquiries for establishing a background for determining obviousness under 35 U.S.C. 103 are summarized as follows: 1. Determining the scope and contents of the prior art. 2. Ascertaining the differences between the prior art and the claims at issue. 3. Resolving the level of ordinary skill in the pertinent art. 4. Considering objective evidence present in the application indicating obviousness or nonobviousness. Claims 1-4, 6-11, 13-18, and 20 are rejected under 35 U.S.C. 103 as being unpatentable over Wallenas et al US Patent Application No. 20170072125 A1 (hereinafter Wallenas) in view of Burbank et al US Patent Application No. 20150005699 A1 (hereinafter Burbank) in view of Mantle US Patent Application No. 20060161107 A1 (hereinafter Mantle). Regarding Claim 1, Wallenas teaches that an apparatus for peritoneal ultrafiltration (i.e., a peritoneal dialysis ("PD") system comprising; Paragraph 0001) is arranged in enclosures (i.e., a housing) designated by broken lines (Fig. 2; Paragraph 0041) that includes a flow pump (i.e., a PD fluid pump housed by the housing; Fig. 2, #65; Paragraph 0075) and two catheter lines (Fig. 2, #53a and 53b) which can be arranged as a dual lumen catheter as is well-known (i.e., a dual lumen patient line extending from the housing; Paragraph 0073) in which the fluid enters the peritoneal cavity through the second catheter (i.e., the dual lumen patient line including a fill lumen Fig. 2, #53b) and out from the first catheter (i.e., and a return lumen; Fig. 2, #53a; Paragraph 0075) with an intermittent tube (i.e., a PD fluid fill line in fluid communication with the fill lumen and an outlet of the PD fluid pump; Fig. 2, #84) and patient line (i.e., connected to a transfer set; a PD fluid return line in fluid communication with the return lumen and an inlet of the PD fluid pump; Fig. 2, #82) connecting the catheters to the apparatus, with pressure sensors arranged in ports (i.e., a fill pressure sensor configured to detect PD fluid pressure along the PD fluid fill line; a return pressure sensor configured to detect PD fluid pressure along the PD fluid return line; Fig. 2, #73 and 86) in order to monitor pressure in the lines (Paragraph 0089) and discusses that a pressure meter may be used for controlling the operation of the flow pump (Paragraph 0093) and that pressure at various points is used to diagnose problems such as occlusions (i.e., to determine if one of the fill lumen or the return lumen of the dual lumen patient line is occluded; Paragraphs 0089-0094). Wallenas further teaches that a partial occlusion would result in a high filling pressure and low emptying pressure which may trigger an alarm (i.e., wherein the control unit is configured to determine (i) an occlusion in the fill lumen if the output from the fill pressure sensor meets or exceeds a first predetermined pressure amount, (ii) an occlusion in the return lumen if the output from the return pressure sensor is less than a second predetermined pressure amount; Paragraphs 0090-0092). Wallenas does not explicitly teach a control unit configured to use outputs from the fill and return pressure sensors and a user interface in operable communication with the control unit, the control unit configured to cause, after determining one of the occlusions in the fill lumen, the return lumen, or the transfer set, the user interface to provide an alarm indicating whether the fill lumen is occluded, the return lumen is occluded, or a transfer set is occluded. However, Burbank teaches a controller (i.e., a control unit; Fig. 1, #116) for regulating flow rate which receives pressure measurements (i.e., configured to use outputs from the fill and return pressure sensors) and applies feedback control to the pump for the purpose of ensuring that the patient’s peritoneal cavity is not over-pressurized (Paragraphs 0055-0057). Burbank further teaches that a user interface is utilized to reference alarms outputs that refer to kinks, obstructions or other types of problems with detailed explanations (i.e., a user interface in operable communication with the control unit, the control unit configured to cause, after determining the occlusion, the user interface to provide an alarm indicating whether the fill lumen is occluded, the return lumen is occluded, or a transfer set is occluded; Paragraph 0162). Burbank is analogous to the claimed invention because it pertains to an automated peritoneal dialysis system (Abstract). It would have been obvious to one of ordinary skill in the art to modify the peritoneal ultrafiltration apparatus as taught by Wallenas with the controller as taught by Burbank because the controller would ensure that the patient’s peritoneal cavity is not over-pressurized. Wallenas in view of Burbank do not explicitly teach (iii) an occlusion in a patient's transfer set if both (i) and (ii) are met. Wallenas teaches multiple pressure sensors and the detection of the individual faults (i) and (ii), but does not explicitly teach the derived fault (iii) when both (i) and (ii) are combined. However, Mantle teaches that it is known, while using a double-barreled catheter for continuous flow dialysis (Paragraph 0025) to include pressure monitoring wherein when pressure is out of control the catheters are tested individually by pressure response to diagnose blockages including blockage of outflow lumen in the OUT and IN directions and the blockage of the inflow lumen in the OUT and IN directions, wherein a blockage of the inflow or outflow of both lumens is non-recoverable and an alarm is signaled (i.e., iii) an occlusion in a patient's transfer set if both (i) and (ii) are met; Paragraph 0070) for the purpose of improving safety by switching the control of the pumps to a manual operation (Paragraph 0069). Mantle is analogous to the claimed invention because it pertains to devices and processes for automated extracorporeal recirculation and chemical manipulation of biological liquids, specifically including liquid pumped to and from the peritoneum (Abstract and Paragraph 0001). It would have been obvious to one of ordinary skill in the art to modify the apparatus for peritoneal ultrafiltration made obvious by Wallenas in view of Burbank with the derived alarm to alert of a blockage of the inflow/outflow lumens of the double-barreled catheter as taught by Mantle because the alert would allow the operator to respond to a non-recoverable position with manual pump operation to improve safety. Regarding Claim 2, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 1. Burbank further teaches that the controller applies feedback control to the pump (i.e., wherein the control unit is further configured to operate the PD fluid pump; Paragraphs 0055-0057) and that low or no flow events trigger alarms (i.e., and to determine if the PD fluid pump is pumping PD fluid or not pumping PD fluid; Paragraph 0161). Regarding Claim 3, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 1. Wallenas further teaches in Fig. 2 that the intermittent line (Fig. 2, #84) and the patient line (Fig. 2, #82) clearly extend into the dotted line enclosure (i.e., wherein the PD fluid fill line is located within the housing). PNG media_image2.png 388 742 media_image2.png Greyscale Regarding Claim 4, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 1. Wallenas further teaches in Fig. 2 that the intermittent line (Fig. 2, #84) and the patient line (Fig. 2, #82) clearly extend into the dotted line enclosure (i.e., wherein the PD fluid return line is located within the housing). PNG media_image2.png 388 742 media_image2.png Greyscale Regarding Claim 6, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 1. Wallenas further teaches that pressure should not be high during standstill of the flow pump and should normally be higher than or equal to atmospheric pressure (i.e., wherein the control unit is configured to compare the outputs from the fill and return pressure sensors when the PD fluid pump is not being actuated, and wherein the control unit determines an absence of an occlusion in the return lumen and the fill lumen when the outputs from the fill and return pressure sensors are the same or almost the same; Paragraphs 0089-0090). Regarding Claim 7, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 1. Wallenas further teaches that a partial occlusion would result in a high filling pressure and low emptying pressure which may trigger an alarm (Paragraph 0091) and that when sudden and unexpected deviation of pressure is measured, an alarm is activated (i.e., wherein the control unit is configured to compare the outputs from the fill and return pressure sensors when the PD fluid pump is being actuated, and wherein the control unit determines an absence of an occlusion in the return lumen and the fill lumen when the outputs from the fill and return pressure sensors are different by a pressure drop amount; Paragraph 0094). Regarding Claim 8, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 7. Wallenas further teaches that the flow pump may continuously circulate peritoneal fluid into and out of the peritoneal cavity by operating the flow pump in the clockwise direction and by opening the intermittent valve (i.e., wherein the PD fluid pump, the PD fluid fill line, the fill lumen, the PD fluid return line and the return lumen form a closed loop; Fig. 2, #68; Paragraph 0075) and that the pressure is measured at the connections to the patient line and intermittent line and the catheters (i.e., and wherein the pressure drop amount includes a pressure drop in both the fill lumen and the return lumen; Paragraph 0089). Regarding Claim 9, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 7. Burbank further teaches a valve (Fig. 19, #P1) located on the drain line (Fig. 19, #862; Paragraph 0151) with automatic control of the valves (a valve under control of the control unit, the valve positioned to close the PD fluid return line; Paragraph 0153). Furthermore, the limitation “wherein the pressure drop amount includes a pressure drop in the fill lumen but not the return lumen while the return lumen is closed by the valve” is a contingent limitation and as long as the structure of the prior art apparatus is capable of performing the function, then the limitation is rendered obvious (Ex parte Schulhauser, Appeal 2013-007847 (PTAB April 28, 2016)). See MPEP 2111.04(II). Regarding Claim 10, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 1. Wallenas further teaches that a partial occlusion would result in a high filling pressure and low emptying pressure which may trigger an alarm (i.e., the output from the fill pressure sensor meets or exceeds the first predetermined pressure amount when the output from the fill pressure sensor is greater than an expected output from the fill pressure sensor by at least a first predetermined amount, and wherein the output from the return pressure sensor is less than a second predetermined pressure amount when the output from the return pressure sensor is less than an expected output from the return pressure sensor by at least a second predetermined amount; Paragraphs 0090-0092). Regarding Claim 11, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 10. Wallenas further teaches that the normal pressures are always above or at atmospheric pressure (i.e., wherein the first and second predetermined amounts are at least substantially the same; Paragraph 0090). Regarding Claim 13, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 1. Wallenas further teaches that the normal pressures are always above or at atmospheric pressure (i.e., wherein the first and second predetermined amounts are at least substantially the same; Paragraph 0090). Regarding Claim 14, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 1. Burbank further teaches fill/drain line tubing with connectors (i.e., connector; Fig. 3B, #326 and 324) at the peritoneal catheter (i.e., at the distal end; Fig. 3B, #322; Paragraph 0064) where variations are known to have separated fill and drain lines, each having a respective lumen that merge at the peritoneal catheter (i.e., wherein the fill lumen and the return lumen of the dual lumen patient line merge into a single lumen in a connector located at the distal end of the dual lumen patient line; Paragraph 0067). Regarding Claim 15, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 1. Burbank further teaches fill/drain line tubing with connectors (i.e., connector; Fig. 3B, #326 and 324) at the peritoneal catheter (i.e., at a distal end; Fig. 3B, #322; Paragraph 0064) where variations are known to have separated fill and drain lines, each having a respective lumen that merge at the peritoneal catheter (i.e., wherein the fill lumen and the return lumen of the dual lumen patient line extend into a connector located at the distal end of the dual lumen patient line; the connector configured to connect to the transfer set for performing continuous flow peritoneal dialysis Paragraph 0067). Regarding Claim 16, Wallenas teaches that an apparatus for peritoneal ultrafiltration (i.e., a peritoneal dialysis ("PD") system comprising; Paragraph 0001) is arranged in enclosures (i.e., a housing) designated by broken lines (Fig. 2; Paragraph 0041) that includes a flow pump (i.e., a PD fluid pump housed by the housing; Fig. 2, #65; Paragraph 0075) and two catheter lines (Fig. 2, #53a and 53b) which can be arranged as a dual lumen catheter as is well-known (i.e., a dual lumen patient line extending from the housing; Paragraph 0073) in which the fluid enters the peritoneal cavity through the second catheter (i.e., the dual lumen patient line including a fill lumen Fig. 2, #53b) and out from the first catheter (i.e., and a return lumen; Fig. 2, #53a; Paragraph 0075) with an intermittent tube (Fig. 2, #84) and patient line (i.e., connected to a transfer set; Fig. 2, #82) connecting the catheters to the apparatus, with pressure sensors arranged in ports (i.e., a fill pressure sensor configured to detect PD fluid pressure in the fill lumen of the dual lumen patient line; a return pressure sensor configured to detect PD fluid pressure in the return lumen of the dual lumen patient line; Fig. 2, #73 and 86) in order to monitor pressure in the lines (Paragraph 0089) and discusses that a pressure meter may be used for controlling the operation of the flow pump (Paragraph 0093) and that pressure at various points is used to diagnose problems such as occlusions (i.e., to determine if one of the fill lumen or the return lumen of the dual lumen patient line is occluded; Paragraphs 0089-0094). Wallenas further teaches that a partial occlusion would result in a high filling pressure and low emptying pressure which may trigger an alarm (i.e., wherein the control unit is configured to determine (i) an occlusion in the fill lumen if the output from the fill pressure sensor meets or exceeds a first predetermined pressure amount, (ii) an occlusion in the return lumen if the output from the return pressure sensor is less than a second predetermined pressure amount; Paragraphs 0090-0092). Wallenas does not explicitly teach a control unit configured to use outputs from the fill and return pressure sensors and a user interface in operable communication with the control unit, the control unit configured to cause, after determining the occlusion, the user interface to provide an alarm indicating whether the fill lumen is occluded, the return lumen is occluded, or a transfer set is occluded. However, Burbank teaches a controller (i.e., a control unit; Fig. 1, #116) for regulating flow rate which receives pressure measurements (i.e., configured to use outputs from the fill and return pressure sensors) and applies feedback control to the pump for the purpose of ensuring that the patient’s peritoneal cavity is not over-pressurized (Paragraphs 0055-0057). Burbank further teaches that a user interface is utilized to reference alarms outputs that refer to kinks, obstructions or other types of problems with detailed explanations (i.e., a user interface in operable communication with the control unit, the control unit configured to cause, after determining one of the occlusions in the fill lumen, the return lumen, or the transfer set, the user interface to provide an alarm indicating whether the fill lumen is occluded, the return lumen is occluded, or a transfer set is occluded; Paragraph 0162). Burbank is analogous to the claimed invention because it pertains to an automated peritoneal dialysis system (Abstract). It would have been obvious to one of ordinary skill in the art to modify the peritoneal ultrafiltration apparatus as taught by Wallenas with the controller as taught by Burbank because the controller would ensure that the patient’s peritoneal cavity is not over-pressurized. Wallenas in view of Burbank do not explicitly teach (iii) an occlusion in a patient's transfer set if both (i) and (ii) are met. Wallenas teaches multiple pressure sensors and the detection of the individual faults (i) and (ii), but does not explicitly teach the derived fault (iii) when both (i) and (ii) are combined. However, Mantle teaches that it is known, while using a double-barreled catheter for continuous flow dialysis (Paragraph 0025) to include pressure monitoring wherein when pressure is out of control the catheters are tested individually by pressure response to diagnose blockages including blockage of outflow lumen in the OUT and IN directions and the blockage of the inflow lumen in the OUT and IN directions, wherein a blockage of the inflow or outflow of both lumens is non-recoverable and an alarm is signaled (i.e., iii) an occlusion in a patient's transfer set if both (i) and (ii) are met; Paragraph 0070) for the purpose of improving safety by switching the control of the pumps to a manual operation (Paragraph 0069). Mantle is analogous to the claimed invention because it pertains to devices and processes for automated extracorporeal recirculation and chemical manipulation of biological liquids, specifically including liquid pumped to and from the peritoneum (Abstract and Paragraph 0001). It would have been obvious to one of ordinary skill in the art to modify the apparatus for peritoneal ultrafiltration made obvious by Wallenas in view of Burbank with the derived alarm to alert of a blockage of the inflow/outflow lumens of the double-barreled catheter as taught by Mantle because the alert would allow the operator to respond to a non-recoverable position with manual pump operation to improve safety. Regarding Claim 17, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 16. Wallenas further teaches that a partial occlusion would result in a high filling pressure and low emptying pressure which may trigger an alarm (i.e., the output from the fill pressure sensor meets or exceeds the first predetermined pressure amount when the output from the fill pressure sensor is greater than an expected output from the fill pressure sensor by at least a first predetermined amount, and wherein the output from the return pressure sensor is less than a second predetermined pressure amount when the output from the return pressure sensor is less than an expected output from the return pressure sensor by at least a second predetermined amount; Paragraphs 0090-0092). Regarding Claim 18, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 17. Wallenas further teaches that the normal pressures are always above or at atmospheric pressure (i.e., wherein the first and second predetermined amounts are at least substantially the same; Paragraph 0090). Regarding Claim 20, Wallenas in view of Burbank in view of Mantle makes obvious the PD system of claim 16. Wallenas further teaches that the normal pressures are always above or at atmospheric pressure (i.e., wherein the first and second predetermined amounts are at least substantially the same; Paragraph 0090). Conclusion Any inquiry concerning this communication or earlier communications from the examiner should be directed to ADAM ADRIEN GERMAIN whose telephone number is (703)756-5499. The examiner can normally be reached Mon - Fri 7:30-4:30. Examiner interviews are available via telephone, in-person, and video conferencing using a USPTO supplied web-based collaboration tool. To schedule an interview, applicant is encouraged to use the USPTO Automated Interview Request (AIR) at http://www.uspto.gov/interviewpractice. If attempts to reach the examiner by telephone are unsuccessful, the examiner’s supervisor, In Suk Bullock can be reached at (571)272-5954. The fax phone number for the organization where this application or proceeding is assigned is 571-273-8300. Information regarding the status of published or unpublished applications may be obtained from Patent Center. Unpublished application information in Patent Center is available to registered users. To file and manage patent submissions in Patent Center, visit: https://patentcenter.uspto.gov. Visit https://www.uspto.gov/patents/apply/patent-center for more information about Patent Center and https://www.uspto.gov/patents/docx for information about filing in DOCX format. For additional questions, contact the Electronic Business Center (EBC) at 866-217-9197 (toll-free). If you would like assistance from a USPTO Customer Service Representative, call 800-786-9199 (IN USA OR CANADA) or 571-272-1000. /A.A.G./ Examiner, Art Unit 1777 /IN SUK C BULLOCK/ Supervisory Patent Examiner, Art Unit 1772
Read full office action

Prosecution Timeline

Oct 31, 2022
Application Filed
Jun 05, 2025
Non-Final Rejection — §103
Sep 09, 2025
Response Filed
Oct 07, 2025
Final Rejection — §103
Jan 16, 2026
Request for Continued Examination
Jan 23, 2026
Response after Non-Final Action
Mar 23, 2026
Non-Final Rejection — §103 (current)

Precedent Cases

Applications granted by this same examiner with similar technology

Patent 12533681
NEW FROTHERS FOR MINERALS RECOVERY
2y 5m to grant Granted Jan 27, 2026
Patent 12303915
USE OF 2-CYANO-N-(SUBSTITUTED CARBAMOYL)ACETAMIDE COMPOUND IN FLOTATION OF CALCIUM-BEARING MINERALS
2y 5m to grant Granted May 20, 2025
Study what changed to get past this examiner. Based on 2 most recent grants.

AI Strategy Recommendation

Get an AI-powered prosecution strategy using examiner precedents, rejection analysis, and claim mapping.
Powered by AI — typically takes 5-10 seconds

Prosecution Projections

3-4
Expected OA Rounds
11%
Grant Probability
-4%
With Interview (-15.0%)
3y 2m
Median Time to Grant
High
PTA Risk
Based on 27 resolved cases by this examiner. Grant probability derived from career allow rate.

Sign in with your work email

Enter your email to receive a magic link. No password needed.

Personal email addresses (Gmail, Yahoo, etc.) are not accepted.

Free tier: 3 strategy analyses per month